Michigan Administrative Code (Last Updated: November 16, 2016) |
![]() |
Department LR. Licensing and Regulatory Affairs |
![]() |
MIOSHA |
![]() |
Chapter Ionizing Radiation Rules Part 1 to Part 15 |
![]() |
Part 7. MEDICAL X-RAY INSTALLATIONS |
Section 325.5321. X-ray equipment.
Latest version.
All data is extracted from pdf, click here to view the pdf.
-
(1) The x-ray equipment shall comply with the general requirements of rule 312 excluding subrules (11), (14) and (16).
(2) Maximum potential shall be mechanically or electronically limited to 85 kVp.
(3) A contact therapy machine shall meet the additional requirement that the leakage radiation at 5 centimeters (2 inches) from the surface of the tube housing shall not exceed 0.1 R/h. As used in this subrule "contact therapy machine" means an x-ray therapy machine designed for source to skin treatment distances of 5 centimeters or less at tube potentials in the range of 20 to 50 kVp.
History: 1954 ACS 85, Eff. Dec. 3, 1975.