Part 7. MEDICAL X-RAY INSTALLATIONS |
Section 325.5353. Conditions of operation.
All data is extracted from pdf, click here to view the pdf.
-
(1) Operation shall comply with the general requirements of rules 333 and
348.
(2) Individuals operating mobile or portable diagnostic x-ray equipment shall wear a
protective apron of minimum 0.5 millimeter lead equivalence unless portable shielding is provided as specified in rule 352 (1).
(3) Mobile or portable diagnostic x-ray equipment shall not be used for routine radiography or fluoroscopy in hospitals or private offices of practitioners of the healing arts. This equipment shall only be used when it is medically inadvisable to move a patient to a fixed radiographic or fixed fluoroscopic installation.