Section 325.5379. X-ray equipment.  


Latest version.

All data is extracted from pdf, click here to view the pdf.

  • (1) X-ray equipment in multiple tube installations shall comply with the general requirements of rule 373 with regard to each tube housing assembly and each complete x-ray unit.

    (2)    When 2 or more x-ray tube heads are operated from a single exposure switch (multiple tube units), there shall be indication at the control panel showing which tube is connected and ready to be  energized,   and  means   to prevent energizing more than 1 tube head at the same time.

    (3)  For multiple tube units there shall be indication at the tube housing assembly when it is connected and ready to be energized.

History: 1979 AC.