Michigan Administrative Code (Last Updated: November 16, 2016) |
Department LR. Licensing and Regulatory Affairs |
MIOSHA |
Chapter Ionizing Radiation Rules Part 1 to Part 15 |
Part 10. VETERINARY X-RAY INSTALLATIONS |
Section 325.5418. Conditions of operation.
All data is extracted from pdf, click here to view the pdf.
-
(1) Each individual present in a fluoroscopic room shall wear a protective apron of at least 0.5 millimeter lead equivalence.
(2) Only individuals whose presence is needed to conduct the examination, to conduct radiation protection surveys or undergoing specific training shall be permitted in the fluoroscopy room during x-ray exposures.
(3) Personnel monitoring shall be performed in controlled areas for each individual occupationally exposed to ionizing radiation from diagnostic x-ray equipment. Personnel monitoring devices such as film badge dosimeters or thermoluminescent dosimeters, shall be permanently assigned to each
occupationally exposed individual. This monitoring shall be continuous during employment as a radiation worker.
(4) Personnel exposure records shall be kept on permanent available file at the facility where the exposure occurs.
(5) Monitoring devices used to estimate whole body exposure shall normally be worn on the chest or abdomen. Monitoring of any other body part shall comply with rule 222. Since employees involved in fluoroscopic procedures are required to wear protective aprons and may be subjected to non-uniform radiation fields, a dosimeter assigned to monitor whole body exposure will not necessarily record the dose most representative of exposure to the lens of the eye. To monitor this critical area for which the exposure limit is the same as for whole body, active blood-forming organs, or gonads, an auxiliary dosimeter shall be provided in accord with rule 222.
(6) Monitoring devices worn to estimate personnel occupational exposure shall not be worn by the individual when he is exposed as a patient for any medical or dental reason.
(7) The fluoroscopist's eyes should be sufficiently dark-adapted for the visual task required before commencing conventional fluoroscopy. Under no circumstances shall he attempt to compensate for inadequate adaptation by increasing exposure factors employed or by prolonging the fluoroscopic examination.
(8) A fluoroscopic x-ray system shall not be left unattended without locking the apparatus, room or building in some manner which will prevent use of the apparatus by unauthorized persons.
History: 1979 AC.