![]() |
Michigan Administrative Code (Last Updated: November 16, 2016) |
![]() |
Department LR. Licensing and Regulatory Affairs |
![]() |
MIOSHA |
![]() |
Chapter Ionizing Radiation Rules Part 1 to Part 15 |
![]() |
Part 13. MISCELLANEOUS SOURCES |
Section 325.5694. Mammography equipment evaluations.
All data is extracted from pdf, click here to view the pdf.
-
(1) Additional evaluations of stereotactic breast biopsy units or image processors shall be conducted whenever a new unit or processor is installed, a unit or processor is disassembled and reassembled at the same or a new location, or major components of a stereotactic breast biopsy unit or processor equipment are changed or repaired. These evaluations shall be used to determine whether the new or changed equipment meets the requirements of R 325.5676 to R 325.5678 and R 325.5687 to R 325.5691, as applicable. Problems revealed by the evaluation shall be corrected before the new or changed equipment is put into service for procedures or film processing.
(2) The equipment evaluations specified in subrule (1) of this rule shall be performed by a medical physicist or by an individual under the direct supervision of a medical physicist.
History: 2013 AACS.